Getting started with the Analog Discovery Studio Max

1. Hardware

To set up the Analog Discovery Studio Max:

  1. First, install the Digilent WaveForms software on your PC. See Installer Details for more information.
  2. Connect the Analog Discovery Studio Max to your PC using a USB cable, 19V DC supply and power switch to ON position.

Recommended operating conditionsAbsolute maximum ratings
Oscilloscopeinputs±50 V 1 MΩ | 15 pF±50 V
Wavegenoutputs±10 V 50Ω 30 mA DC±5.8 V
DigitalIOs0 V / 3.3 V 8 mA±15 V
TriggerIO0 V / 3.3 V 8 mA±15 V
Positive Supplyoutput1 V 15 V 500 mA
Negative Supplyoutput-1 V -15 V 500 mA
PCI Suppliesoutput3.3 V, 5 V, +15 V, -15 V
Device Supplyinput19 V DC 4.74 A90 W 18.2 V .. 19.8 V

For more information visit the Resource Center.

 


2. Software

See Installer Details for information about different command line options to install WaveForms.
See WaveForms for more information on using the WaveForms software.

 


3. Troubleshooting

If the device is not detected by WaveForms application or experience random disconnects usually indicates contact problem. In this case, try using the original USB cable the device or other USB cables you have at hand, try using other USB port and hub. Make sure the 19V DC supply is plugged in properly.

 


4. Oscilloscope

The Analog Discovery Studio Max has 4 oscilloscope input channels.

Specifications:

The coupling and other options can be selected under channel options in Scope, Spectrum, Network and Impedance Analyzers.

 


5. Arbitrary Waveform Generator

The Analog Discovery Studio Max is equipped with two Arbitrary Waveform Generators channels.

Specifications:

 


6. Digital I/O

The Analog Discovery Studio Max has 24 digital I/Os (DIO 0-23).

Specifications:

 


7. Power Supplies

The Analog Discovery Studio Max has two variable power supplies.

Specifications:

Output Voltage+1 V to +15 V-1 V to -15 V
Current Limit20 mA to 500 mA-20 mA to -500 mA
Voltage Accuracy100 mV100 mV
Ripple and Noise20 mV pk2pk20 mV pk2pk
Voltage Readback Accuracy±15 mV±15 mV
Current Readback Accuracy±5 mV±5 mV

These power supplies can be used as slow AWG channels, see Wavegen channel 3 and 4.
Soft start is adjustable as full scale time between 1ms and 84.5ms, default is 66ms.

The supply rails on the PCI connector are:

Output Voltage3.3 V5 V+15 V-15 V
Accuracy±5 %±5 %±5 %±5 %
Current Output310 mA2 A500 mA500 mA

For more information visit the Resource Center.

 


8. Digital Multimeter

The Analog Discovery Studio Max has a DMM with AC or DC voltage or current, resistance, voltage drop across diode, continuity.

Specifications:

Resolution4.5 digits
Input impedance10 MΩ
Voltage input protection±60 V
Current input protection 4.5 A fuse, 5MF 3.5-R
Voltage measurement
DC ranges50 mV, 500 mV, 5 V, 50 V DC
AC ranges50 mV, 500 mV, 5 V, 30 V RMS
AC input frequency range40 Hz to 8 kHz
Accuracy0.2% of 50 mV range and 0.1% at other ranges*
Current measurement
Range2 A DC/RMS
Shunt resistance20 mΩ
AC input frequency range40 Hz to 8 kHz
Accuracy0.1% or range*
Resistance measurement
Ranges50 Ω, 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 50 MΩ
Accuracy1% of 50 Ω and 50 MΩ ranges, 0.1% of other ranges

* at 50 Hz for AC